Reprocessing A to D and the 1 May 2028 deadline

Open in the coursewith narrated slides, a checklist to tick off and a quiz

At a glance

In detail

The four levels

Annex 2 of the hygiene regulation orders reprocessing by the treatment an instrument was used for (ÖZÄK-HygV 2025, Annex 2):

LevelContact withExamplesReprocessing
Aintact skin and mucosaexamining the mouthcleaning and disinfection, by machine or by hand with listed products; visual check; dust-free storage
Bmucosa, enamel, dentinefillings, preparation, crowns, scaling, prostheticscleaning; disinfection by machine, or instead in the steriliser (class B autoclave); visual check; dust-free storage
Cpulp; penetrating the mucosaroot canal treatment, extraction, periodontal surgerycleaning and disinfection by machine or a validated manual method; visual check; sterilised in sterile packaging; dust-free storage for at most 6 months
Doral surgeryimplants, sinus lift, flap surgerycleaning and disinfection by machine; visual check; sterile-packed in a class B steriliser; dust-free storage for at most 6 months

What the regulation requires

The deadline

The requirements of § 20(4) and (5) must be met by 1 May 2028 at the latest § 24(3) HygV; everything else in §§ 20 to 22 has applied since 1 May 2025 § 24(1) HygV. The chamber describes this as a three-year transition for buying a class B steriliser in practices that already existed on 1 May 2025 (Dental Chamber, hygiene).

Does every practice need a washer-disinfector? No. § 20(4) allows manual or machine methods, depending on the intended use. A washer-disinfector is needed by anyone reprocessing level D instruments; for level B, the class B autoclave can replace machine disinfection, and for level C a validated manual method is enough (Annex 2). A class B steriliser, on the other hand, is needed by every practice that sterilises.

Checklist

Quiz

What must be in place by 1 May 2028 at the latest under § 24(3) HygV?

  1. The requirements for cleaning and disinfection and for the steriliser (§ 20(4) and (5))
  2. An annual external validation
  3. A separate sterilisation room
  4. A washer-disinfector in every practice
Show the answer

The answer is A: The requirements for cleaning and disinfection and for the steriliser (§ 20(4) and (5)). The requirements of § 20(4) and (5) must be met by 1 May 2028 at the latest (§ 24(3) ÖZÄK-HygV 2025). Paragraph 4 allows manual or machine cleaning depending on the intended use; a washer-disinfector isn't required in every practice.

Sources

This lesson's statements rest on:

Not legal advice. What counts is the acts and regulations in the Federal Legal Information System and the Austrian Dental Chamber's regulations (read on 7 October 2026). The health minister's regulation with the criteria from 2028 (§ 9b(2) GQG) had not been issued on that day. Not covered are insurer contracts, billing, data protection in detail and the employer's duties under the Worker Protection Act in detail; there is a separate course on those at aschg.globalapps.top. Not an offer of the Austrian Dental Chamber, a regional dental chamber, the BIQG or the Chamber of Commerce.