Dental practice course All courses

Lesson 4.3 · 5 min

Safety checks and the electrical inspection

Open in the coursewith narrated slides, a checklist to tick off and a quiz

At a glance

In detail

The safety check

The operator has active non-implantable medical devices safety-checked regularly, and non-active ones only if the manufacturer requires it § 6(1) MPBV. Purely battery-powered devices not in Annex 1 are exempt, unless the manufacturer requires checks § 6(2) MPBV.

QuestionAnswer
How often?at the manufacturer's interval; if there is none, a qualified person sets it: as a rule 6 to 36 months, for Annex 1 devices 6 to 24 months. An interval set this way may be overrun by up to six months§ 6(3), (5)
Who?only people or bodies with technical training, testing experience and measuring equipment under Annex 3§ 6(7)
What stays?a report with the inspector, date, scope, measured values and tolerances, methods and overall result, kept for at least five years§ 6(8)
On the device?a label with the date of the next check (month, year) and who did the check§ 6(9)

Older labels under the 2007 version stay valid until the end of their check interval § 12(7) MPBV. Annex 1 devices get an incoming inspection before first use; if a detailed test report from the manufacturer or supplier comes with it, a visual check for transport damage and the label are enough § 3 MPBV.

The electrical installation

Anyone who employs staff is covered by the Electrical Protection Regulation 2012 § 1 ESV:

The QSV requires that "required inspections take place as prescribed" § 5 no. 5 QSV. On one sample visit, the electrical inspection report had to be renewed (quality report 2017–2021).

Checklist

Quiz

How often must a residual current device be checked with the test button if the manufacturer gives no interval?

  1. Only after a power cut
  2. Every day
  3. Every five years
  4. At least every six months
Show the answer

The answer is D: At least every six months. Residual current devices are checked by pressing the test button at the manufacturer's intervals, or if none are given, at least every six months and after a fault (§ 7(3) no. 1 ESV 2012).

Sources

This lesson's statements rest on:

Not legal advice. What counts is the acts and regulations in the Federal Legal Information System and the Austrian Dental Chamber's regulations (read on 7 October 2026). The health minister's regulation with the criteria from 2028 (§ 9b(2) GQG) had not been issued on that day. Not covered are insurer contracts, billing, data protection in detail and the employer's duties under the Worker Protection Act in detail; there is a separate course on those at aschg.globalapps.top. Not an offer of the Austrian Dental Chamber, a regional dental chamber, the BIQG or the Chamber of Commerce.